Sign in →

Test Code LAB908 Hepatitis B Virus e Antigen, Serum

Additional Codes

EAG

Useful For

Determining the presence or absence of detectable hepatitis B virus e antigen in monitoring infection status of individuals with chronic hepatitis B

 

Determining infectivity of hepatitis B virus (HBV) carriers

 

Monitoring serologic response of chronically HBV-infected patients receiving antiviral therapy

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum SST


Ordering Guidance


If ordered with HBVQN / Hepatitis B Virus (HBV) DNA Detection and Quantification by Real-Time PCR, Serum; send separate vials.



Additional Testing Requirements


 



Necessary Information


Date of collection is required.



Specimen Required


Patient Preparation: For 24 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).

Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Serum gel (red-top tubes are not acceptable)

Submission Container/Tube: Plastic vial

Specimen Volume: 0.7 mL

Collection Instructions:

1. Centrifuge blood collection tube per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).

2. Aliquot serum into plastic vial.


Specimen Minimum Volume

0.6 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum SST Frozen (preferred) 90 days
  Refrigerated  6 days
  Ambient  72 hours

Reference Values

Negative

Day(s) Performed

Monday through Saturday

CPT Code Information

87350

LOINC Code Information

Test ID Test Order Name Order LOINC Value
EAG Hepatitis Be Ag, S 13954-3

 

Result ID Test Result Name Result LOINC Value
EAG Hepatitis Be Ag, S 13954-3

Interpretation

Presence of hepatitis B virus e antigen (HBeAg) and absence of HBe antibody (anti-HBe) usually indicate active hepatitis B virus (HBV) replication and high infectivity.

 

Absence of HBeAg with appearance of anti-HBe is consistent with loss of HBV infectivity.

Method Description

The Elecsys HBeAg (hepatitis B e antigen) assay is based on the sandwich principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. HBeAg present in patient's sample reacts with 2 biotinylated monoclonal anti-HBeAg antibodies and a mixture of monoclonal anti-HBeAg antibody and polyclonal anti-HBeAg antibodies labeled with a ruthenium complex react to form a sandwich complex. After addition of streptavidin-coated microparticles (solid phase), the complexes bind to this solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode, which induces chemiluminescent emission that is measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the patient’s sample to the cutoff index value set from reagent lot-specific assay calibration.(Package insert: Elecsys HBeAg. Roche Diagnostics; v1.0, 10/2020)

Report Available

1 to 3 days

Specimen Retention Time

14 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject
Heat-inactivated specimen Reject

Method Name

Electrochemiluminescence Immunoassay (ECLIA)

Forms

If not ordering electronically, complete, print, and send 1 of the following:

-Gastroenterology and Hepatology Test Request (T728)

-Infectious Disease Serology Test Request (T916)

Secondary ID

80510