Test Code LAB796 Hepatitis B Virus e Antibody, Serum

Additional Codes

HEAB

Useful For

Determining the presence or absence of detectable hepatitis B virus e antibody in monitoring infection status of individuals with chronic hepatitis B

 

Determining infectivity of hepatitis B virus (HBV) carriers

 

Monitoring serologic response of chronically HBV-infected patients receiving antiviral therapy

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum SST


Additional Testing Requirements


If ordered with HBVQN / Hepatitis B Virus (HBV) DNA Detection and Quantification by Real-Time PCR, Serum; send separate vials.



Necessary Information


Date of collection is required.



Specimen Required


Patient Preparation: For 24 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).

Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Serum gel (red-top tubes are not acceptable)

Submission Container/Tube: Plastic vial

Specimen Volume: 0.7 mL

Collection Instructions:

1. Centrifuge blood collection tube per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).

2. Aliquot serum into plastic vial.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum SST Frozen (preferred) 90 days
  Refrigerated  6 days
  Ambient  72 hours

Reference Values

Negative

 

See Viral Hepatitis Serologic Profiles.

Day(s) Performed

Monday through Saturday

CPT Code Information

86707

LOINC Code Information

Test ID Test Order Name Order LOINC Value
HEAB HBe Antibody, S 33463-1

 

Result ID Test Result Name Result LOINC Value
HEAB HBe Antibody, S 33463-1

Interpretation

Absence of hepatitis B e antigen (HBeAg) with appearance of HBe antibody (anti-HBe) is consistent with inactivity of the virus and loss of hepatitis B virus (HBV) infectivity.

 

Although resolution of chronic HBV infection generally follows the appearance of anti-HBe, the HBV carrier state may persist.

Method Description

The Elecsys Anti-HBe (hepatitis B virus e antibody) assay is based on the competitive immunoassay principle and performed using an electrochemiluminescence method on the automated cobas e 801 immunochemistry analyzer. Anti-HBe present in the patient's sample binds to the added synthetic HBe antigen (HBeAg). The remaining unbound sites on the synthetic HBeAg become occupied with the added biotinylated antibodies and ruthenium complex-labeled antibodies specific for HBeAg. The entire complex becomes bound to streptavidin-coated microparticles (solid phase) via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. After unbound substances are washed away, voltage is applied to the electrode, which induces chemiluminescent emission that is measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the reaction product to the cutoff index value set from reagent lot-specific assay calibration.(Package insert: Elecsys Anti-HBe. Roche Diagnostics; v1.0, 12/2021)

Report Available

Same day/1 to 3 days

Specimen Retention Time

14 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject

Method Name

Electrochemiluminescence Immunoassay (ECLIA)

Secondary ID

80973

Forms

If not ordering electronically, complete, print, and send 1 of the following with the specimen:

-Infectious Disease Serology Test Request (T916)

-Gastroenterology and Hepatology Test Request (T728)