Test Code LAB7490 JC Virus, Molecular Detection, PCR, Spinal Fluid
Additional Codes
LCJC
Reporting Name
JC Virus PCR, CSFUseful For
Aids in diagnosing progressive multifocal leukoencephalopathy due to JC virus (JCV)
This test is not to be used as a diagnostic tool for Creutzfeldt-Jakob disease (CJD).
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
CSFSpecimen Required
Specimen Required
Supplies: Sarstedt 5 mL Aliquot Tube (T914)
Preferred: 12 x 75-mm screw cap vial (T465)
Acceptable: Sterile screw cap vial
Container/Tube: Sterile vial
Specimen Volume: 0.5 mL
Collection Instructions: Do not centrifuge.
Specimen Minimum Volume
0.3 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
CSF | Refrigerated (preferred) | 7 days | |
Frozen | 7 days |
On-campus collections: Tube to 99 or deliver to University Extension Hospital Room EH318
Off-campus collections: Do NOT Spin! Specimen to be stored/transported at Refrigerated temp.
Reference Values
Negative
Day(s) Performed
Monday through Friday
Test Classification
This test was developed, and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
87798
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
LCJC | JC Virus PCR, CSF | 33295-7 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
88909 | JC Virus PCR, CSF | 33295-7 |
Clinical Information
JC virus (JCV), a member of the genus Polyomavirus, is a small nonenveloped DNA-containing virus. Primary infection occurs in early childhood, with a prevalence of greater than 80%.(1) The virus is latent but can reactivate in immunosuppressed patients, especially those with AIDS.
JCV is recognized as the etiologic agent of progressive multifocal leukoencephalopathy (PML), a fatal demyelinating disease of the central nervous system.(2,3) Histologic examination of brain biopsy tissue may reveal characteristic pathologic changes localized mainly in oligodendrocytes and astrocytes. Detection of JCV DNA by PCR (target gene, large T antigen) in the cerebrospinal fluid specimens of patients with suspected PML infection has replaced the need for biopsy tissue for laboratory diagnosis.(4) Importantly, the PCR test is specific with no cross-reaction with BK virus, a closely related polyomavirus.
Interpretation
Detection of JC virus (JCV) DNA supports the clinical diagnosis of progressive multifocal leukoencephalopathy due to JCV.
Cautions
A negative result does not rule out the possibility of JC virus (JCV) infection.
The reference range in cerebrospinal fluid is "negative" for this assay, although JCV DNA may be detectable in the absence of clinical symptoms in certain patient populations.(5,6) However, this assay is only to be used for patients with appropriate neurological and neuroradiological features of progressive multifocal leukoencephalopathy, and is not indicated for screening asymptomatic patients.
Supportive Data
The following data supports the use of this assay for clinical testing.
Accuracy:
Twenty-six negative cerebrospinal fluid (CSF) specimens were spiked with JC virus (JCV)-positive control plasmid at the limit of detection (approximately 10 targets/mcL). The 26 spiked specimens were run in a blinded manner with 14 negative (nonspiked) specimens. 100% of the spiked specimens were positive and 100% of the nonspiked specimens were negative.
Analytical Sensitivity/Limit of Detection (LoD:
The lower limit of detection (LoD) of this assay is 10 DNA target copies per mcL in CSF.
Analytical Specificity:
No PCR signal was obtained from the extracts of 15 viral isolates that may cause similar symptoms or be found in the CSF, including herpes simples virus (HSV) types 1 and 2, Epstein-Barr virus (EBV), cytomegalovirus (CMV), human herpesvirus (HHV)-6, HHV-7, HHV-8, enterovirus, mumps, adenovirus, BK virus, and Simian virus 40 (SV40).
Precision:
Interassay precision was 100% and intraassay precision was 100%.
Reference Range:
The reference range in CSF is "negative" for this assay.
Reportable Range:
This is a qualitative assay and the results are reported as either negative or positive for targeted JCV DNA.
Clinical Reference
1. Safak M, Khalili K: An overview: human polyomavirus JC virus and its associated disorders. J Neurovirol 2006;9 Suppl 1:3-9
2. Khalili K, White MK: Human demyelinating disease and the polyomavirus JCV. Mult Scler 2006 Apr;12(2):133-142
3. Ahsan N, Shah KV: Polyomaviruses and human diseases. Adv Exp Med Biol 2006;577:1-18
4. Romero JR, Kimberlin DW: Molecular diagnosis of viral infections of the central nervous system. Clin Lab Med 2003 Dec;23(4):843-865
5. Chen Y, Bord E, Tompkins T, et al: Asymptomatic reactivation of JC virus in patients treated with natalizumab. N Engl J Med 2009 Sep 10;361(11):1067-1074
6. Egli A, Infanti L, Dumoulin A, et al: Prevalence of polyomavirus BK and JC infection and replication in 400 healthy donors. J Infect Dis 2009 Mar 15;199(6):837-846
Method Description
Viral nucleic acid is extracted from the specimen using the MagNA Pure automated instrument (Roche Applied Science). Primers are directed to the large T antigen gene, which is a conserved sequence specific for JC virus (JCV). This assay detects only JCV; it does not detect BK virus or Simian virus 40 (SV40) (other polyomaviruses). The LightCycler instrument (Roche Applied Science) amplifies and monitors the development of target nucleic acid sequences after the annealing step during PCR cycling. This automated PCR system can rapidly detect amplicon development through stringent air-controlled temperature cycling in capillary cuvettes. The detection of amplified products is based on the fluorescence resonance energy transfer (FRET) principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3'-end is excited by an external light source and emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, at the 5'-end. The acceptor fluorophore then emits a light of a different wavelength that can be measured with a signal that is proportional to the amount of specific PCR product.(Unpublished Mayo method)
Report Available
2 to 5 daysSpecimen Retention Time
1 weekReject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability. |
NY State Approved
YesMethod Name
Real-Time Polymerase Chain Reaction (PCR)/DNA Probe Hybridization
Forms
If not ordering electronically, complete, print, and send a Microbiology Test Request (T244) with the specimen.
Testing Algorithm
See Meningitis/Encephalitis Panel Algorithm in Special Instructions.