Test Code LAB2846 Hepatitis B Virus Surface Antibody, Qualitative/Quantitative, Serum
Additional Codes
HBAB
Useful For
Identifying previous exposure to hepatitis B virus
Determining adequate immunity from hepatitis B vaccination
Testing Algorithm
For more information see Hepatitis B: Testing Algorithm for Screening, Diagnosis, and Management
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
Serum SSTOrdering Guidance
If patient is being monitored for hepatitis B immune globulin (HBIG) therapy after organ transplantation, order HBABT / Hepatitis B Virus Surface Antibody Monitor, Post-Transplant, Serum.
This test should not be used for screening asymptomatic, nonpregnant individuals with or without risk factors for hepatitis B virus (HBV) infection. For screening such patients, order HBBSN / Hepatitis B Virus Surface Antibody Screen, Qualitative/Quantitative, Serum.
This test should not be used for prenatal screening of pregnant individuals with or without risk factors for HBV infection. For screening such patients, order HBABP / Hepatitis B Virus Surface Antibody Prenatal, Qualitative/Quantitative, Serum.
Necessary Information
Date of collection is required.
Specimen Required
Patient Preparation: For 24 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Serum gel (red-top tubes are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 0.7 mL
Collection Instructions:
1. Centrifuge blood collection tube per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).
2. Aliquot serum into a plastic vial.
Specimen Minimum Volume
0.6 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum SST | Frozen (preferred) | 90 days | |
Refrigerated | 6 days | ||
Ambient | 7 hours |
Special Instructions
Reference Values
Hepatitis B Surface Antibody
Unvaccinated: Negative
Vaccinated: Positive
Hepatitis B Surface Antibody, Quantitative
Unvaccinated: <8.5 mIU/mL
Vaccinated: ≥11.5 mIU/mL
Day(s) Performed
Monday through Saturday
CPT Code Information
86706
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
HBAB | HBs Antibody, S | 5193-8 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
HB_AB | HBs Antibody, S | 10900-9 |
HBSQN | HBs Antibody, Quantitative, S | 5193-8 |
Interpretation
A positive result indicates recovery from acute or chronic hepatitis B virus (HBV) infection or acquired immunity from HBV vaccination. This assay does not differentiate between a vaccine-induced immune response and an immune response induced by HBV. A positive hepatitis B core (anti-HBc) total antibody result would indicate that the hepatitis B surface antibody (anti-HBs) response is due to past HBV infection.
Per assay manufacturer's instructions for use, positive results, defined as anti-HBs levels of 11.5 mIU/mL or greater, indicate adequate immunity to HBV from past hepatitis B or HBV vaccination. However, per current Centers for Disease Control and Prevention guidance,(1) individuals with anti-HBs levels greater than 10 mIU/mL after completing an HBV vaccination series are considered protected from hepatitis B infection.
Negative results, defined as anti-HBs levels of less than 8.5 mIU/mL, indicate a lack of recovery from acute or chronic hepatitis B or inadequate immune response to HBV vaccination. The US Advisory Committee on Immunization Practices does not recommend more than 2 HBV vaccine series in vaccine nonresponders.
Indeterminate results, defined as anti-HBs levels in the range from 8.5 to less than 11.5 mIU/mL, indicate inability to determine if anti-HBs is present at levels consistent with recovery or immunity. Repeat testing is recommended in 1 to 2 months to determine definitive anti-HBs status.
For more information see Hepatitis B: Testing Algorithm for Screening, Diagnosis, and Management.
Method Description
The Elecsys Anti-HBs (hepatitis B virus surface antibody) assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescent immunoassay on the automated cobas e 801 immunochemistry analyzer. Anti-HBs present in patient's serum sample reacts with the biotinylated HBs antigen (ad and ay subtypes) and HBs antigen (ad/ay) labeled with a ruthenium complex to form a sandwich complex. After the addition of streptavidin-coated microparticles (solid phase), the complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode, which induces chemiluminescent emissions that is measured by a photomultiplier. The emission signal generated is directly proportional to the concentration of anti-HBs present in the patient's sample. (Package insert: Elecsys Anti-HBs. Roche Diagnostics; v3.0, 03/2024)
Report Available
Same day/1 to 3 daysSpecimen Retention Time
14 daysReject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Heat-inactivated specimen | Reject |
Method Name
Electrochemiluminescence Immunoassay (ECLIA)
Forms
If not ordering electronically, complete, print, and send 1 of the following: