Test Code LAB1879 Human Herpesvirus-6, Molecular Detection, PCR, Plasma

Additional Codes

HHV6

Useful For

As an adjunct in the rapid diagnosis of human herpesvirus-6 infection using plasma specimens

 

This test should not be used to screen asymptomatic patients.

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Plasma EDTA


Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Lavender top (EDTA)

Submission Container/Tube:

Preferred: Plastic vial

Acceptable: Screw-capped, sterile container

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot plasma into plastic vial.


Specimen Minimum Volume

0.3 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Plasma EDTA Refrigerated (preferred) 7 days
  Frozen  7 days

Reference Values

Negative

Day(s) Performed

Monday through Saturday

CPT Code Information

87532

LOINC Code Information

Test ID Test Order Name Order LOINC Value
HHV6 HHV-6 PCR, P 29495-9

 

Result ID Test Result Name Result LOINC Value
87532 HHV-6 PCR, P 29495-9

Interpretation

A positive result indicates the presence of specific DNA from human herpesvirus-6 (HHV-6) and supports the diagnosis of infection with this virus.

 

A negative result indicates the absence of detectable DNA from HHV-6 in the specimen, but it does not negate the presence of the virus or active or recent disease.

Method Description

Viral nucleic acid is extracted by the MagNA Pure automated instrument (Roche Applied Science) from clinical specimens. Primers directed to the immediate early gene of human herpesvirus-6 (HHV-6) produce a 195-base pairs amplicon. The LightCycler instrument (Roche Applied Science) amplifies and monitors the development of target nucleic acid sequences after the annealing step during polymerase chain reaction (PCR) cycling. This automated PCR system can rapidly (1 hour) detect amplicon development through stringent air-controlled temperature cycling and capillary cuvettes. The detection of amplified products is based on the fluorescence resonance energy transfer (FRET) principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3'-end is excited by an external light source and emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, at the 5'-end. The acceptor fluorophore then emits light of a different wavelength that can be measured with a signal that is proportional to the amount of specific PCR product. LightCycler hybridization probes are designed for exact homology to HHV-6.(Cockerill FR III, Uhl JR: Applications and challenges of real-time PCR for the clinical microbiology laboratory. In: Reischel U, Wittwer C, Cockerill F. Rapid Cycle Real-Time PCR - Methods and Applications. Springer-Verlag;2002: 3-30)

Report Available

2 to 5 days

Specimen Retention Time

1 week

Reject Due To

Gross hemolysis Reject
Gross lipemia OK

Method Name

Real-Time Polymerase Chain Reaction (PCR)/DNA Probe Hybridization

Forms

If not ordering electronically, complete, print, and send a Microbiology Test Request (T244) with the specimen.

Secondary ID

87532