Test Code GAPCR Streptococcus Group A, Molecular Detection, PCR, Throat
Specimen Required
Patient Preparation: Do not collect throat swab specimens immediately after use of antiseptic mouthwash.
Specimen Type: Throat swab
Supplies: BD E-Swab (T853)
Container/Tube:
Preferred: BD Liquid Amies Elution Swab (ESwab)
Acceptable: Copan Liquid Amies Elution Swab (ESwab)
Specimen Volume: Entire collection/1 ESwab
Collection Instructions:
1. Must be collected and transported using the ESwab collection kit.
2. Using a tongue blade to hold the tongue down, firmly rub the swab over the tonsils, posterior pharynx, and any areas of inflammation with exudate. Ensure that areas with exudate or inflammation are swabbed.
3. Avoid the tongue, teeth, and gums as the swab is withdrawn.
4. Insert the swab back into the ESwab transport tube and break the swab at the indicated score line.
5. Label specimen according to policy.
6. Send refrigerated. Do not freeze.
Secondary ID
610254Useful For
Detecting group A streptococcal infections in patients with signs and symptoms of pharyngitis using a throat swab specimen
The test is not intended for monitoring treatment for group A Streptococcus infections.
Method Name
Real-Time Polymerase Chain Reaction (PCR)
Specimen Type
VariesSpecimen Minimum Volume
See Specimen Required
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Varies | Refrigerated (preferred) | 6 days | |
Ambient | 48 hours |
Reject Due To
Specimens collected in media other than ESwab collection kit | Reject |
Reference Values
Negative
Interpretation
Positive: Streptococcus pyogenes DNA is detected.
Negative: Streptococcus pyogenes DNA is not detected.
Method Description
The GeneXpert Dx System automates and integrates sample purification, nucleic acid extraction and amplification, and detection of the target sequence using real-time polymerase chain reaction (PCR). The system requires single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, cross-contamination between samples is minimized.
The Xpert Xpress Strep A test, performed on the GeneXpert Instrument Systems, is a rapid, qualitative test using real-time PCR to detect Streptococcus pyogenes DNA. The assay includes a sample processing control to ensure an adequate amplification process and to monitor for the presence of inhibitors in the PCR reaction. A probe check control verifies reagent rehydration, PCR tube filling, and all reaction components, including probes and dyes, are present and functional in the cartridge.(Package insert: Xpert Xpress Strep A. Cepheid; 301-9326, Rev. D, 06/2019)
Day(s) Performed
Monday through Sunday
Report Available
Same day/1 daySpecimen Retention Time
6 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterCPT Code Information
87651
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
GAPCR | Group A Streptococcus PCR, Throat | 60489-2 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
GAPCR | Group A Streptococcus PCR, Throat | 60489-2 |