Test Code CMVMS Cytomegalovirus Antibody, IgM, Serum


Specimen Required


Supplies: Sarstedt Aliquot Tube 5 mL (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 0.6 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Secondary ID

621834

Useful For

Aiding in the diagnosis of acute infection with cytomegalovirus

Method Name

Electrochemiluminescence Immunoassay (ECLIA)

Specimen Type

Serum

Specimen Minimum Volume

See Specimen Required

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 28 days
  Frozen  90 days
  Ambient  7 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject
Additives (eg, biocides, antioxidants) Reject
Heat inactivated Reject

Reference Values

Negative

 

Reference values apply to all ages.

Interpretation

Negative:

No IgM to cytomegalovirus (CMV) detected. False negative results may occur in immunocompromised patients.

 

Borderline:

Recommend follow-up testing in 10 to 14 days if clinically indicated.

 

Positive:

CMV IgM antibodies detected, which may indicate active or recent infection. Low level IgM antibodies may persist for more than 12 months following disease resolution.

Method Description

The electrochemiluminescence immunoassay for the in vitro qualitative determination of IgM antibodies to cytomegalovirus (CMV) in human serum is a micro-capture test principle. During the first incubation, 6 mcL of sample are automatically prediluted 1:20 with Diluent Universal. Biotinylated monoclonal anti-h-IgM-specific antibodies are added. The second incubation, CMV-specific recombinant antigen labeled with a ruthenium complex and streptavidin-coated microparticles are added. The complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell II M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration.(Package insert: Elecsys CMV IgM. Roche Diagnostics GmbH; 11/2022)

Day(s) Performed

Monday through Saturday

Report Available

Same day/1 to 3 days

Specimen Retention Time

14 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

86645

LOINC Code Information

Test ID Test Order Name Order LOINC Value
CMVMS CMV Ab, IgM, S 30325-5

 

Result ID Test Result Name Result LOINC Value
MCMV CMV Ab, IgM, S 30325-5