Test Code ADMB ADAMTS13 Inhibitor Bethesda Titer, Plasma
Specimen Required
Only orderable as a reflex. For more information see ADAMP / ADAMTS13 Activity with Reflex Inhibitor Profile, Plasma.
Specimen Type: Platelet-poor plasma
Patient Preparation: Fasting preferred
Collection Container/Tube: Light-blue top (3.2% sodium citrate)
Submission Container/Tube: Plastic vials
Specimen Volume: 2 mL in 2 plastic vials, each containing 1 mL
Collection Instructions:
1. Specimen must be collected prior to factor replacement therapy.
2. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing.
3. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.
4. Aliquot plasma (1 mL per aliquot) into 2 separate plastic vials leaving 0.25 mL in the bottom of centrifuged vial.
5. Freeze plasma immediately (no longer than 4 hours after collection) at -20° C or ideally, at or below -40° C.
Additional Information:
1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.
2. Each coagulation assay requested should have its own vial.
Specimen Stability Information: Frozen 2 years
Secondary ID
620817Useful For
Assisting with the diagnosis of congenital, immune or acquired thrombotic thrombocytopenic purpura as a part of a profile
Special Instructions
Method Name
Only orderable as a reflex. For more information see ADAMP / ADAMTS13 Activity with Reflex Inhibitor Profile, Plasma.
Fluorescence Resonance Energy Transfer (FRET)
Specimen Type
Plasma Na CitSpecimen Minimum Volume
2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma Na Cit | Frozen |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Reference Values
Only orderable as a reflex. For more information see ADAMP / ADAMTS13 Activity with Reflex Inhibitor Profile, Plasma.
<0.5 BU
Interpretation
Assisting with the diagnosis of congenital, immune or acquired thrombotic thrombocytopenic purpura as a part of a profile
Method Description
The ADAMTS13 activity is measured by a fluorescence resonance energy transfer-based assay using a synthetic fragment of von Willebrand factor as substrate. Cleavage of this small fragment by the ADAMTS13 protease generates fluorescence that is directly proportionate to the quantification of ADAMTS13 activity. The Bethesda titer assay is performed using mixing studies. One inhibitor (Bethesda) unit is defined as the concentration of an inhibitor that is able to reduce ADAMTS13 activity of normal pooled plasma by 50%.(Package insert: ATS-13 ADAMTS13 Activity Assay 2.0. Immucor; 08/2023)
Day(s) Performed
Monday through Friday, Sunday
Report Available
1 to 3 daysSpecimen Retention Time
14 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterCPT Code Information
85335
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
ADMB | ADAMTS13 Inhibitor Titer, P | 40824-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
620817 | ADAMTS13 Inhibitor Titer | 40824-5 |